Publikasi Scopus 926 artikel (Per 14 Maret 2022)

Lubis A.C., Iqbal M., Munawar D.A., Hartono B., Munawar M.
57208338816;57194421966;56470745000;46961152700;16747447600;
A simple percutaneous retrieval technique for an embolized watchman left atrial appendage closure device in the thoracic aorta using a homemade snare: A case report
2021
International Heart Journal
62
5
1153
1155
Binawaluya Cardiac Center, Jakarta, Indonesia; Faculty of Medicine, Universitas Sumatera Utara, Sumatera Utara, Indonesia; Department of Cardiology and Vascular Medicine, Medical Faculty, Universitas Indonesia, Jakarta, Indonesia
Lubis, A.C., Binawaluya Cardiac Center, Jakarta, Indonesia, Faculty of Medicine, Universitas Sumatera Utara, Sumatera Utara, Indonesia; Iqbal, M., Binawaluya Cardiac Center, Jakarta, Indonesia; Munawar, D.A., Binawaluya Cardiac Center, Jakarta, Indonesia, Department of Cardiology and Vascular Medicine, Medical Faculty, Universitas Indonesia, Jakarta, Indonesia; Hartono, B., Binawaluya Cardiac Center, Jakarta, Indonesia; Munawar, M., Binawaluya Cardiac Center, Jakarta, Indonesia, Department of Cardiology and Vascular Medicine, Medical Faculty, Universitas Indonesia, Jakarta, Indonesia
A 60-year old male with paroxysmal atrial fibrillation underwent a combined procedure of left atrial appendage occlusion and pulmonary vein isolation. However, an acute intraprocedural 24-mm Watchman device dislodgement occurred, and thus a decision for urgent surgery was made. However, it was noted during the surgery that the device had migrated further to the descending aorta, just distal to the left subclavian artery. Since a right sternotomy access for retrieval was not feasible, a percutaneous approach was justified. A homemade snare was created using a combination of a long sheath, J-wire, and a regular snare, and the device was successfully retrieved without significant difficulty. © 2021, International Heart Journal Association, All rights reserved.
Atrial fibrillation; Complication; Left atrial appendage occlusion
anticoagulant agent; adult; anticoagulation; Article; case report; clinical article; descending aorta; device embolization; drug withdrawal; fluoroscopy; general anesthesia; human; left subclavian artery; male; middle aged; occlusion; paroxysmal atrial fibrillation; percutaneous drainage; percutaneous retrieval technique; pulmonary vein; pulmonary vein isolation; sternotomy; surgical technique; thoracic aorta; transesophageal echocardiography; vascular access; adverse device effect; atrial fibrillation; device removal; devices; heart atrium appendage; heart catheterization; heart surgery; multimodality cancer therapy; procedures; septal occluder; subclavian artery; thoracic aorta; treatment outcome; Aorta, Thoracic; Atrial Appendage; Atrial Fibrillation; Cardiac Catheterization; Cardiac Su
International Heart Journal Association
13492365
34544965
Article
Q2
555
9100